For each of NIF's 192 beams:

  • The pulse shape as a function of time must be generated with a high degree of precision
  • The energy delivered to the target must be precise
  • The energy must be delivered to the target at exactly the specified time

NIF's injection laser system (ILS) plays a key role in meeting these three requirements. The ILS system is responsible for generating a prescribed pulse shape, adjusting the energy in each of the 192 beams, and adjusting the time it takes the beams to hit the target. The ILS consists of the master oscillator room (MOR), the preamplifier module (PAM), the input sensor package (ISP), and the preamplifier beam transport system (PABTS). NIF has one master oscillator, 48 PAMs and ISPs, and 24 PABTS's.

Master Oscillator Room
NIF laser pulses are generated in the MOR from an optical fiber laser. This laser generates pulses of just a few nanojoules (billionths of a joule) with a beam diameter of a few micrometers.

Master Oscillator Room equipment generates the low-energy laser pulses that are subsequently amplified several billion times before being injected into the NIF laser amplifiers.

The pulse is then split and amplified using fiber splitters and fiber amplifiers similar to those used in the telecommunications industry. After several stages of splitting and amplification, 48 individual pulses are generated. Each pulse is then shaped in time per the experimental requirements and transported to each of the 48 PAMs for further amplification and beam shaping.

Preamplifier Module
The PAM contains two different amplification stages. The first amplifies the pulse energy from about 750 picojoules (trillionths of a joule) to about 15 millijoules (thousandths of a joule), a factor of 20 million. The second stage amplifies the shaped beam by a factor of up to 10,000. The output energy of the PAM can be varied from about 10 millijoules to about 10 joules.

Input Sensor Package
The ISP performs two functions—aligning the PAM and main laser optical system, and measuring the PAM output energy, beam profile, and pulse shape. Because there are no other energy adjustment devices in the main laser chain, the PAM energy must be precisely adjusted so that each beam on the ignition target has the requested energy. A fast photodiode measures the pulse shape in time to make sure that the pulses have the proper shape.

Preamplifier Beam Transport System
The pair of beams produced from the two PAMs is split into four beams in the PABTS system using a series of beam splitters (each beam is split in two and then the two beams are each split into two again). The PABTS can adjust the split ratio—the percentage of light into each of the four beams. In addition, the PABTS contains a series of mirrors for each beam that can be moved to change the distance each beam propagates in the PABTS. This in turn changes the amount of time it takes for each beam to reach the ignition target.